How to switch from MALDI to ESI mode - WKB241374
Article number: 241374
OBJECTIVE or GOAL
Switch from MALDI to ESI mode.
ENVIRONMENT
- MALDI SYNAPT G2-Si
PROCEDURE
- First put the instrument into Standby and vent. Wait 15 to 30 minutes for the vacuum to reach atmosphere.
- Disconnect the source ID cable(s) from the instrument panel.
- Release the inlet clamps as shown below.
- The Inlet should slide back away from the source. Lower it via the key pad on the right-hand side.
- Remove the two MALDI plates:
- Confirm that the O-rings are seated in the PEEK block, and fit the ESI ion block.
- Fit the ESI source enclosure and set the instrument to pump down. This may take several hours.
ADDITIONAL INFORMATION
id241374, Electrospray, MALDIG2SI, MALDIG2SI, MALDIMICRO, SYNG2SIHD