Fluidics valve always leads to waste on the TQD, Xevo MS Absolute, Xevo MS, Xevo TQD, Xevo TQ-S, Xevo TQ-S micro or Xevo TQ-XS - WKB236184
Article number: 236184
SYMPTOMS
- Fluidics valve always goes to waste
- Nitrogen generator 100 psi
- Source pressure test successful
- No error message on the Instrument Console
ENVIRONMENT
- Xevo TQ-S micro
- TQD
- Xevo TQ MS
- Xevo TQ-S
- Xevo TQ-XS
- Xevo TQ Absolute
CAUSE
API gas is OFF.
FIX or WORKAROUND
Turn on the API gas.
ADDITIONAL INFORMATION
This message is normal whenever the API gas is shut off, the instrument is shut down, or the MS is rebooted. It is a built-in safety mechanism to divert the flow to waste when the API gas is off so the source does not flood from LC flow.