Intermittent missing peaks on Xevo TQ-XS - WKB225349

Article number: 225349

SYMPTOMS

  • Intermittent missing peaks on Xevo TQ-XS
  • LC working as expected
  • Injection volume and vial types are correct
  • Method events are enabled 
  • Electronic reboot did not resolve the issue

ENVIRONMENT

  • Xevo TQ-S
  • Xevo TQ-XS
  • MassLynx 4.1
  • MassLynx 4.2

CAUSE

Method events are not working as intended, inhibiting the diverter valve function

FIX or WORKAROUND

  1. Create a new method with at least one MRM starting from 0.00 minutes.
  2. An acquisition window must be present at or slightly before the time of the event.