Skip to main content
Waters

Intermittent missing peaks on Xevo TQ-XS - WKB225349

Article number: 225349

SYMPTOMS

  • Intermittent missing peaks on Xevo TQ-XS
  • LC working as expected
  • Injection volume and vial types are correct
  • Method events are enabled 
  • Electronic reboot did not resolve the issue

ENVIRONMENT

  • Xevo TQ-S
  • Xevo TQ-XS
  • MassLynx 4.1
  • MassLynx 4.2

CAUSE

Method events are not working as intended, inhibiting the diverter valve function

FIX or WORKAROUND

  1. Create a new method with at least one MRM starting from 0.00 minutes.
  2. An acquisition window must be present at or slightly before the time of the event.

ADDITIONAL INFORMATION

Refer to WKB21632.

id225349, MLYNX, MLYNXV41, SUPMM, XEVOG2XSQT, XEVOTQS, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD

Not able to find a solution? Click here to request help.