Intermittent missing peaks on Xevo TQ-XS - WKB225349
Article number: 225349
SYMPTOMS
- Intermittent missing peaks on Xevo TQ-XS
- LC working as expected
- Injection volume and vial types are correct
- Method events are enabled
- Electronic reboot did not resolve the issue
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ-XS
- MassLynx 4.1
- MassLynx 4.2
CAUSE
Method events are not working as intended, inhibiting the diverter valve function
FIX or WORKAROUND
- Create a new method with at least one MRM starting from 0.00 minutes.
- An acquisition window must be present at or slightly before the time of the event.
ADDITIONAL INFORMATION
Refer to WKB21632.
id225349, MLYNX, MLYNXV41, SUPMM, XEVOG2XSQT, XEVOTQS, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD