Skip to main content
Waters

Low sensitivity with Xevo QTof system - WKB225524

Article number: 225524

SYMPTOMS

  • Low sensitivity in all modes of operation

ENVIRONMENT

  • Xevo QTof
  • Xevo G2
  • Synapt G2-Si

CAUSE

The detector voltage is not optimized.

FIX or WORKAROUND

  1. Infuse leucine enkephaline solution.
  2. Run detector setup in IntellliStart.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.