Scroll pump not turning fully on to start the pump down of the SYNAPT G2-S - WKB20811
Article number: 20811
SYMPTOMS
- Scroll vibrates as though it is turned on, but never comes to full speed.
- Mass spec times out and cannot pump down.
ENVIRONMENT
- Xevo TQ-MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
- SYNAPT HDMS
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- Xevo QTof
- Xevo Tof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-XS
CAUSE
The scroll pump is not receiving the full 220V from the UPS. The read back is 88V and 110V with a volt meter. The wall outlet where the UPS is plugged in reads 110V and 110V.
FIX or WORKAROUND
Customer must service or replace the UPS.