Source pressure test fails on the Xevo TQ MS, Xevo TQD, Xevo TQ-S, Xevo TQ-S cronos, Xevo TQ-S micro or Xevo TQ-XS - WKB208398
Article number: 208398
SYMPTOMS
- System shows leak warning when performing source enclosure pressure diagnosis, and then fails the pressure test
ENVIRONMENT
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
The ferrule on the waste trap assembly tubing connection from the MS is damaged.
FIX or WORKAROUND
Replace the trap bottle PN 700004606, or PN 700005811 for TQS Micro.
ADDITIONAL INFORMATION