Skip to main content
Waters

Source pressure test fails on the Xevo TQ MS, Xevo TQD, Xevo TQ-S, Xevo TQ-S cronos, Xevo TQ-S micro or Xevo TQ-XS - WKB208398

Article number: 208398

SYMPTOMS

  • System shows leak warning when performing source enclosure pressure diagnosis, and then fails the pressure test

ENVIRONMENT

  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

The ferrule on the waste trap assembly tubing connection from the MS is damaged.

FIX or WORKAROUND

Replace the trap bottle PN 700004606, or PN 700005811 for TQS Micro.

ADDITIONAL INFORMATION

 

id208398, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQGC, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD

Not able to find a solution? Click here to request help.