IntelliStart Sample Tune and Develop Method (STDM) fails to find daughter ion, even though a strong signal is seen in the Tune page and with manual acquisition from the Tune page - WKB208055

Article number: 208055

SYMPTOMS

  • IntelliStart fails to find daughter ion

ENVIRONMENT

  • MassLynx
  • Empower 3
  • IntelliStart Sample Tune and Develop Method (STDM)

CAUSE

IntelliStart is set to exclude any masses below 100 AMU.

FIX or WORKAROUND

Change the daughter ion exclusion mass to a value lower than the possible daughter ion.