Xevo TQ-XS low sensitivity post PM - WKB205509
Article number: 205509
SYMPTOMS
- Loss in sensitivity (both positive and negative modes) post PM.
ENVIRONMENT
- Xevo TQ-XS
- Xevo TQD
- TQ Detector
- Xevo TQS micro
- Xevo TQS cronos
- Xevo TQ Absolute
CAUSE
- Capillary protrusion changed during PM servicing of the ESI probe.
FIX or WORKAROUND
- Adjust the capillary protrusion at a flow rate the user normally runs at to optimize sensitivity.
- User should be shown how to optimize the capillary position as most never adjust this.
- You can find the optimum lengh ( 0.5-1mm out of the probetip ) inside the usermanual from the MS.
ADDITIONAL INFORMATION
