"Source door open" error with Xevo mass spectrometer - WKB205519
Article number: 205519
SYMPTOMS
- "Source door open" error
- "Low desolvation gas" error
- Loss of communication between mass spectrometer and PC
ENVIRONMENT
- Xevo TQ-S micro
- Xevo TQ-S
- Xevo TQD
- Xevo TQ-MS
- Xevo TQ-S cronos
- Xevo TQ-XS
- Xevo G2-XS
CAUSE
Faulty Ethernet cable connecting mass spectrometer to computer or UPLC instrument
FIX or WORKAROUND
Replace the Ethernet cable.
ADDITIONAL INFORMATION