Leak on MS divert valve (fluidics diverter valve blocked) - WKB195369
Article number: 195369
SYMPTOMS
- Probe is not blocked when bypassing valve
- No beam on infusion of setup solution
- Fittings leak at diverter valve despite tightening
ENVIRONMENT
- Xevo TQ-XS
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-S cronos
- Xevo TQD
CAUSE
Blockage in diverter valve
FIX or WORKAROUND
- Remove the front of the valve (stator) and the black insert (rotor) and sonicate in methanol for 15 minutes. Examine both stator and rotor for signs of wear.
- If wear is observed or there is still a leak at the diverter valve, replace both using the Diverter Rotor/Stator Kit (Part Number 700009464) for Xevo TQ-XS and TQ-S.
- For Xevo TQ-S micro, Xevo TQ-S cronos and Xevo TQD order Diverter Rotor/Stator Kit (Part Number 700009465).