How do I power-off or shut down a QDa detector that has been running? - WKB19525
Article number: 19525
OBJECTIVE or GOAL
Power-off or shut down a QDa detector.
ENVIRONMENT
- ACQUITY QDa Mass Detector
- Empower
- MassLynx
PROCEDURE
- Stop the mobile phase system flow to the QDa detector.
- Take the detector out of Operate mode and put it into Standby mode using the control software.
- Turn off the QDa using its power button.
- The system should automatically turn off the nitrogen gas and vent in about five minutes.
ADDITIONAL INFORMATION
id19525, eluent, QDA