APGC failed to start second run in sample list queue - WKB19550
Article number: 19550
SYMPTOMS
- APGC fails to start second run in sample list.
- MS waiting for injection to start.
ENVIRONMENT
- MassLynx
- Xevo G2 QTof
- APGC
- HP7890
- HP7683B
CAUSE
GC configuration not set correctly.

FIX or WORKAROUND
Set the split/splitless injector parameters to "Ignore Ready" False.
- On the GC, select CONFIG > (Back or Front) Inlet > Enter.
- Go to Ignore Ready and select [Mod/Type] to adjust to FALSE.