Xevo TQD calibration file expired - WKB17232

Article number: 17232

SYMPTOMS

  • Console gives a Calibration File Expired error message.
  • The Xevo TQD was calibrated the day before.

ENVIRONMENT

  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQD
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • SQD2

CAUSE

Wrong resolution mode selected.

FIX or WORKAROUND

  1. Go to IntelliStart in the console.
  2. Under the Type of Analysis mode drop-down menu, select a calibrated resolution mode.

XTQD select res mode.PNG

ADDITIONAL INFORMATION

id17232, A-SQ, SQD2, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD