Calibration file expired on the SQD2, Xevo TQ MS, Xevo TQD, Xevo TQ-S, Xevo TQ-S Micro or Xevo TQ-MS - WKB17232

Article number: 17232

SYMPTOMS

  • Console gives a Calibration File Expired error message.
  • The Xevo TQD was calibrated the day before.

ENVIRONMENT

  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQD
  • Xevo TQ-S Micro
  • Xevo TQ-XS
  • SQD2

CAUSE

Wrong resolution mode selected.

FIX or WORKAROUND

  1. Go to IntelliStart in the console.
  2. Under the Type of Analysis mode drop-down menu, select a calibrated resolution mode.

XTQD select res mode.PNG

ADDITIONAL INFORMATION