Vials are lifted out of sample tray as I-Class needle withdraws - WKB17240
Article number: 17240
SYMPTOMS
- Vials are being lifted out of sample tray as needle withdraws
- Cap mats are lifted from the well plates
ENVIRONMENT
- ACQUITY UPLC Sample Manager Common Platform
CAUSE
In some positions the needle is not penetrating the pre-slit part of the mat or vial. The mat or vial then sticks to the needle.
FIX or WORKAROUND
- Calibrate the rotary tray
- Calibrate the R axis
- Calibrate the needle z axis
ADDITIONAL INFORMATION
id17240, A-30SM, UPISMFTN, UPPFTN, UPSMFTN, UPSMFTN+