Vials are lifted out of sample tray as I-Class needle withdraws - WKB17240

Article number: 17240

SYMPTOMS

  • Vials are being lifted out of sample tray as needle withdraws
  • Cap mats are lifted from the well plates

ENVIRONMENT

  • ACQUITY UPLC Sample Manager Common Platform

CAUSE

In some positions the needle is not penetrating the pre-slit part of the mat or vial. The mat or vial then sticks to the needle.

FIX or WORKAROUND

  1. Calibrate the rotary tray
  2. Calibrate the R axis
  3. Calibrate the needle z axis

ADDITIONAL INFORMATION

 