Do I need to calibrate my quadrupole MS weekly? - WKB15571
Article number: 15571
ENVIRONMENT
- MassLynx
- SQ Detector
- SQ Detector 2
- TQ Detector
- Xevo TQD
- Xevo TQ MS
- Xevo TQ-S
- Xevo TQ-XS
- Xevo TQ-S micro
ANSWER
The customer's standard operating procedure (SOP) stipulates how often the instruments need to be calibrated. The MassLynx online help states the following:
Recommendation: Calibrate the mass scale approximately every 3 to 6 months, or run a suitable calibrant to determine mass accuracy. Calibrate before you use methods outside the speed or mass range covered by your current calibration.
ADDITIONAL INFORMATION