Error with Xevo G2-XS: MS Not ready. Detector service due - WKB15638

Article number: 15638

SYMPTOMS

  • Can override this error and run

ENVIRONMENT

  • MassLynx V4.1
  • Masslynx V4.2
  • Xevo G2-XS
  • Xevo G3

CAUSE

Detector setup has not been run in over a month.

FIX or WORKAROUND

  1. Run the detector setup.

ADDITIONAL INFORMATION

It is best to follow the detector setup with the following:

  • Resolution Optimization (three or four times)
  • Calibration
  • LockSpray setup with the LC on