High vacuum error for ion guide region - WKB1543
Article number: 1543
SYMPTOMS
- System status error: High vacuum error for Ion guide region
- Error came after source maintenance
- Ion guide Pirani readback = 3.5 e-3 (normally = 5 e-5)
- With isolation valve closed, pressure drops to under 9e-5
- With isolation valve open, pressure rises back to 4 e-3
ENVIRONMENT
- Xevo TQ-S
CAUSE
Damaged sample cone O-ring
FIX or WORKAROUND
- Replace the sample cone O-ring (700004321).
ADDITIONAL INFORMATION
