Leak detected error message on the SQD, SQD2, TQD, Xevo TQ MS, Xevo TQD, Xevo TQ-S cronos, Xevo TQ-S Micro or Xevo TQ-XS - WKB14692
Article number: 14692
SYMPTOMS
- Xevo TQD or XEVO TQ-S micro, error reported in console: Leak detected
ENVIRONMENT
- ACQUITY SQD system
- ACQUITY TQD system
- TQ Detector
- SQ Detector 2
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
- Blocked capillary or blocked PEEK tubing from valve to ESI probe is causing leak at MS valve port S. Most likely, PEEK tubing is coming loose from PEEK connector due to back pressure from blocked capillary or blocked PEEK tubing.
FIX or WORKAROUND
- Open the front panel by flipping the panel up.
- At the right side of the probe, inspect the drip tray of the MS valve for fluid.

- Disconnect the PEEK tubing from ESI probe. If the leak persists, the PEEK tubing is blocked. Replace PEEK tubing.
- If the PEEK tubing is ok, most likely, the capillary is blocked.
- Replace the capillary.
- Clean the leak sensor.
- Start the flow and inspect for leaks.
ADDITIONAL INFORMATION
See TQD user manual, page 91, for instructions on how to replace the capillary.
P/N 700011241 (standard 500 mm probe); 700011242 (750 mm probe).