TargetLynx is integrating the wrong peak - WKB14747
Article number: 14747
SYMPTOMS
- Peak of interest is not being captured by TargetLynx. A different (incorrect) peak is often selected.
ENVIRONMENT
- TargetLynx
CAUSE
"Update Method Times" is selected.
FIX or WORKAROUND
- In TargetLynx, go to Process Samples.
- Within Process Samples, clear the "Update Method Times" selection.
ADDITIONAL INFORMATION
"Update Method Times" tells TargetLynx to go after whatever peak it thinks should be the target of interest. If there is a peak of greater intensity or better shape than the peak of interest, this incorrect peak is often selected.