MassLynx Method Editor error: The auto dwell time calculated is different from the last saved dwell time. Press Yes to load the calculated dwell time or press No to retain the saved dwell time - WKB121007

Article number: 121007

SYMPTOMS

  • Error message upon opening the MassLynx MS method: The auto dwell time calculated is different from the last saved dwell time. Press Yes to load the calculated dwell time or press No to retain the saved dwell time
  • No chromatographic signal

ENVIRONMENT

  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ MS
  • Xevo TQD
  • SQD2
  • Xevo TQ-XS
  • Xevo TQ-S cronos

CAUSE

The system is set to use manual settings for Inter-scan Delay on the tune page.

clipboard_e0a00d7846f3c8dcc7996d8ec7c540d64.png

clipboard_e2db0983d08709e91c2157ac71cba35f9.png

FIX or WORKAROUND

  1. Set to Automatic Acquisition Inter-scan Delay Settings.
  2. Clear the "Enable Manual Mode" check box in the Inter-scan setup on the tune page.