Skip to main content
Waters

QDa does not collect data but shows "Running" in control panel and does not stop after run time is complete - WKB120905

Article number: 120905

SYMPTOMS

  • Upon injection, QDa does not collect data but shows "Running" in control panel
  • UV detector on the same instrument collects data as normal
  • If run is stopped or run time completes, QDa does not return to Ready state without being reset
  • C drive on Empower PC shows bitlocker encryption error state (exclamation point in a yellow triangle)

ENVIRONMENT

  • QDa
  • Empower

CAUSE

Bitlocker encryption on the C drive is in a paused/error state does not allow full communication between Empower and QDa.

FIX or WORKAROUND

  1. Right-click Bitlocker entry on Taskbar and select "Resume Protection".
  2. The exclamation point warning on the C drive will disappear and the QDa will collect data as normal.

ADDITIONAL INFORMATION

Note that Bitlocker encryption does not need to be disabled; it just needs to not be in an error state.

Not able to find a solution? Click here to request help.