Low sensitivity on ACQUITY QDa - WKB120724
Article number: 120724
SYMPTOMS
- Low area counts on SIR analysis
- Calibration is valid
ENVIRONMENT
- ACQUITY QDa Mass Detector
CAUSE
Ion source components require cleaning.
FIX or WORKAROUND
- Clean the sample cone and ion block in accordance with the source maintenance instructions found in the Overview and Maintenance Guide.
- Run samples to confirm that system sensitivity is as expected.
ADDITIONAL INFORMATION
Follow the appropriate section of the video on www.waters.com
Note that the video follows the RADIAN ASAP procedures, but removing, disassembling, cleaning, and refitting the source components is identical to the QDa Detector.
Removal and disassembly of source components 00:02:34 - 00:06:52 mins
Cleaning source components 00:06:52 - 00:11:12 mins
Refitting source components 00:11:12 - 00:15:23 mins
id120724, QDA