Low sensitivity on ACQUITY QDa - WKB120724

Article number: 120724

SYMPTOMS

  • Low area counts on SIR analysis
  • Calibration is valid

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

Ion source components require cleaning.

FIX or WORKAROUND

  1. Clean the sample cone and ion block in accordance with the source maintenance instructions found in the Overview and Maintenance Guide.
  2. Run samples to confirm that system sensitivity is as expected.

ADDITIONAL INFORMATION

Follow the appropriate section of the video on www.waters.com
Note that the video follows the RADIAN ASAP procedures, but removing, disassembling, cleaning, and refitting the source components is identical to the QDa Detector.

Removal and disassembly of source components 00:02:34 - 00:06:52 mins

Cleaning source components 00:06:52 - 00:11:12 mins

Refitting source components 00:11:12 - 00:15:23 mins