Tubing pops out for ESI probe when flow starts - WKB118473

Article number: 118473

SYMPTOMS

  • Leakage seen at the MS divert valve
  • Tubing pops out for either LC inlet or ESI probe when flow starts

ENVIRONMENT

  • SQ Detector 2
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S Cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • Xevo TQ Absolute
  • Xevo G2-S QToF
  • Xevo G2-XS QToF
  • Xevo G3 QToF
  • Synapt G2-Si
  • Synapt XS

CAUSE

Blocked capillary

FIX or WORKAROUND

Rebuild the probe:
Part number for ESI Probe Spares Kit = 700004292

OR

Exchange one-piece probe:
ESI PROBE ASSEMBLY 500LG x 125uM with part number 700011241

ESI PROBE ASSEMBLY 750LG x 125uM with part number 700011242

ADDITIONAL INFORMATION

id118473, A-SQ, Electrospray, MALDIG2SI, SQD2, SYNAPTXS, SYNG2SIHD, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSTF, XEVOG3QTOF, XEVOTQA, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVTQSMIVD