Tubing pops out for ESI probe when flow starts - WKB118473
Article number: 118473
SYMPTOMS
- Leakage seen at the MS divert valve
- Tubing pops out for either LC inlet or ESI probe when flow starts
ENVIRONMENT
- SQ Detector 2
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S Cronos
- Xevo TQ-S micro
- Xevo TQ-XS
- Xevo TQ Absolute
- Xevo G2-S QToF
- Xevo G2-XS QToF
- Xevo G3 QToF
- Synapt G2-Si
- Synapt XS
CAUSE
Blocked capillary
FIX or WORKAROUND
Rebuild the probe:
Part number for ESI Probe Spares Kit = 700004292
OR
Exchange one-piece probe:
ESI PROBE ASSEMBLY 500LG x 125uM with part number 700011241
ESI PROBE ASSEMBLY 750LG x 125uM with part number 700011242
ADDITIONAL INFORMATION