Article number: 118586

SYMPTOMS

  • "Probe not detected error" appears in the console

ENVIRONMENT

  • Xevo TQ-S micro
  • SQD

CAUSE

Poor connection on the front end enclosure/probe

FIX or WORKAROUND

  1. Reseat the probe and probe cable.
  2. Reseat the desolvation heater cable.
  3. Open and close the source enclosure.

ADDITIONAL INFORMATION