Skip to main content
Waters

Waters Generic Instrument COM Server error in MassLynx - WKB115259

Article number: 115259

SYMPTOMS

  • Sample manager collects sample, moves to the injection port, and then the sample list pauses
  • "Waters Generic Instrument COM Server. Detector Method Failed to open raw files, PreStartdataAcquisition failed" error from the inlet editor

ENVIRONMENT

  • MassLynx 4.1
  • 2767 Sample Manager

CAUSE

Corrupt project folder

FIX or WORKAROUND

  1. Create a new project, method, and sample list
  2. Delete the system and reconfigure it in the new project.

ADDITIONAL INFORMATION

See WKB93646 and WKB2038 for other "Generic Instrument COM server errors.

 

id115259, 2767, 2767SFC, MLYNX, MLYNXV41, SUPMM

Not able to find a solution? Click here to request help.