Waters Generic Instrument COM Server error in MassLynx - WKB115259

Article number: 115259

SYMPTOMS

  • Sample manager collects sample, moves to the injection port, and then the sample list pauses
  • "Waters Generic Instrument COM Server. Detector Method Failed to open raw files, PreStartdataAcquisition failed" error from the inlet editor

ENVIRONMENT

  • MassLynx 4.1
  • 2767 Sample Manager

CAUSE

Corrupt project folder

FIX or WORKAROUND

  1. Create a new project, method, and sample list
  2. Delete the system and reconfigure it in the new project.

ADDITIONAL INFORMATION

See WKB93646 and WKB2038 for other "Generic Instrument COM server errors.