Waters Generic Instrument COM Server error in MassLynx - WKB115259
Article number: 115259
SYMPTOMS
- Sample manager collects sample, moves to the injection port, and then the sample list pauses
- "Waters Generic Instrument COM Server. Detector Method Failed to open raw files, PreStartdataAcquisition failed" error from the inlet editor
ENVIRONMENT
- MassLynx 4.1
- 2767 Sample Manager
CAUSE
Corrupt project folder
FIX or WORKAROUND
- Create a new project, method, and sample list
- Delete the system and reconfigure it in the new project.