Error: Waters Generic Instrument COM Server. Detector Method EnableRun failed with error: 80010105 in Inlet Method - WKB2038
Article number: 2038
SYMPTOMS
- Sample List acquisition paused
- Inlet Method Error window:
Waters Generic Instrument COM Server.
Detector Method EnableRun failed with error: 80010105
The server presented an exception. - Also received the following error: ACQUITY PDA Detector Server has stopped working
ENVIRONMENT
- MassLynx
- ACQUITY PDA Detector
- 2998 Photodiode Array Detector
CAUSE
Unknown
FIX or WORKAROUND
- Deconfigure the inlet modules in the Inlet Editor *.
- Remove entries from the DHCP Server Wizard.
- Restart the PC.
- Configure the Inlet Editor.
- Restart instruments.
ADDITIONAL INFORMATION
* Open Inlet Method Editor > Tools > Instrument Configuration > Configure > Next - you are advised to take screenshots of the configuration before setting to "None"