Error: Waters Generic Instrument COM Server. Detector Method EnableRun failed with error: 80010105 in Inlet Method - WKB2038

Article number: 2038

SYMPTOMS

  • Sample List acquisition paused
  • Inlet Method Error window:
    Waters Generic Instrument COM Server.
    Detector Method EnableRun failed with error: 80010105
    The server presented an exception.
  • Also received the following error: ACQUITY PDA Detector Server has stopped working

ENVIRONMENT

  • Windows 7
  • MassLynx
  • ACQUITY PDA Detector

CAUSE

Not yet known.

FIX or WORKAROUND

  1. Deconfigure the inlet modules in the Inlet Editor.
  2. Remove entries from the DHCP Server Wizard.
  3. Restart the PC.
  4. Configure the Inlet Editor.
  5. Restart instruments.