Poor QTof resolution on intact proteins or monoclonal antibodies (mAb) - WKB115197
Article number: 115197
SYMPTOMS
- Poor resolution on intact proteins or monoclonal antibodies
ENVIRONMENT
- Xevo QTof
- Xevo G2 QTof
- Xevo G2-S
- Xevo G2-XS
- Synapt HDMS
- SYNAPT G2-S
- SYNAPT G2-Si
- Vion IMS QTof
- ACQUITY RDa
- BioAccord
CAUSE
Salt residues in source - The presence of Na or K creates adducts, resulting in a complicated spectrum.
FIX or WORKAROUND
Replace the sample cone with a new one.
ADDITIONAL INFORMATION
As an alternative action, increasing the cone gas flow rate may reduce the negative effect on the intact protein envelop. See the related KCS article WKB117940.
id115197, MALDIG2SI, RDA, SYNG2SIHD, VION, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOQTOF