How is the gradient curve (composition) set when the curve is set in the inlet method? - WKB99596
Article number: 99596
ENVIRONMENT
- MassLynx
- Empower
- Alliance HPLC System
- ACQUITY UPLC
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class
ANSWER
%B = IC + (FC – IC)(t/T) m
IC: Initial condition
FC: Final condition (in the gradient interval)
T: End time of the gradient interval
t: Gradient elapsed time
m: Curve number 6 = X 5 = X1/2
7 = X2 4 = X1/3
8 = X3 3 = X1/4
9 = X4 2 = X1/5
10 = X5 (1 and 11 are stepwise)