ACQUITY UPLC leak error - WKB99542

Article number: 99542

SYMPTOMS

  • Leak error message on console
  • No peaks during sample run
  • Leaking from the inlet port of the flow cell

ENVIRONMENT

  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC Tunable UV Detector

CAUSE

Detector inlet fitting with ferrule out of position 

clipboard_e8ad0ac4036a15a2a085490d13e6318c2.png

clipboard_e272236a8c276a7fd9953ae154a3e28c1.png

FIX or WORKAROUND

  1. Replace or correct the detector inlet fitting. 
  2. Check for back pressure on flow cell or detector outlet.
  3. Flush the system to confirm no leak.
  4. If leak is still present, back flush the flow cell.

ADDITIONAL INFORMATION

For part numbers of UPLC PEEK fittings/assemblies, see the article What is the part number for the inlet tubing to an ACQUITY PDA detector? - WKB10850

If the inlet fitting is still leaking, cut off about an inch of the PEEK tubing and install a new ferrule. 