Sample Fluidics High Pressure limit error on an ACQUITY Sample Manager - WKB99602
Article number: 99602
SYMPTOMS
- Sample Fluidics High Pressure limit error during injection
- Sample Fluidics High Pressure limit error when syringes are purging
ENVIRONMENT
- ACQUITY UPLC Sample Manager
- AQCUITY UPLC I-Class SM-FL
CAUSE
Position 3 of the injection valve cartridge is blocked
FIX or WORKAROUND
- Remove the needle from position 3 of the injection valve cartridge.
- Prime the sample syringe several times to remove the blockage and reduce the sample pressure. Open Console > Sample Manager> Control> Prime Syringe
- If the blockage can't be removed, then replace the Sample Manager Injection Valve part number #700002765 for classic SM-FL or #700006069 for I-class SM-FL.
ADDITIONAL INFORMATION
Error not only seen on blocked port 3, also 1 of the other ports can be blocked and give the same error.