SM-FTN error "Leak detected" - WKB95362
Article number: 95362
SYMPTOMS
- Error message "Leak detected"
- No flow
ENVIRONMENT
- ACQUITY APC Sample Manager FTN
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC I-Class
- ACQUITY UPLC I-Class PLUS
CAUSE
Liquid pooled in the leak sensor port.
FIX or WORKAROUND
- Wipe off any liquid adhering to the leak sensor and remove the liquid in the leak sensor fitting port.
- Leak sensors are built into the upper and lower right-hand sides of the FTN. Check both.
- Reset the device.
ADDITIONAL INFORMATION
id95362, A-30SM, UPAQSM, UPARBQSM, UPARBSMFTN, UPARBSMFTN, UPARCH30, UPARHQSM, UPARHSMFTN, UPARPFTN, UPARQSM, UPARSMFTN, UPARSMFTN, UPBSMIVD, UPIBSM, UPIBSM+, UPIBSMIVD, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPPDAARCB, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMIVD, UPSMPMGR, UPSMPMGR