SM-FTN error "Leak detected" - WKB95362

Article number: 95362

SYMPTOMS

  • Error message "Leak detected"
  • No flow

ENVIRONMENT

  • ACQUITY APC Sample Manager FTN
  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS

CAUSE

Liquid pooled in the leak sensor port.

FIX or WORKAROUND

  1. Wipe off any liquid adhering to the leak sensor and remove the liquid in the leak sensor fitting port.
  2. Leak sensors are built into the upper and lower right-hand sides of the FTN. Check both.
  3. Reset the device.

ADDITIONAL INFORMATION

See "Responding to a leak-sensor alarm" in the ACQUITY UPLC SM-FTN Overview and Maintenance Information (715005051).