SM-FTN error "Leak detected" - WKB95362
Article number: 95362
SYMPTOMS
- Error message "Leak detected"
- No flow
ENVIRONMENT
- ACQUITY APC Sample Manager FTN
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC I-Class
- ACQUITY UPLC I-Class PLUS
CAUSE
Liquid pooled in the leak sensor port.
FIX or WORKAROUND
- Wipe off any liquid adhering to the leak sensor and remove the liquid in the leak sensor fitting port.
- Leak sensors are built into the upper and lower right-hand sides of the FTN. Check both.
- Reset the device.
ADDITIONAL INFORMATION