Skip to main content
Waters

Saturated peaks on FLR detector - WKB94394

Article number: 94394

SYMPTOMS

  • Saturated peaks on the chromatogram

ENVIRONMENT

  • ACQUITY UPLC FLR Detector
  • 2475 Multi-Wavelength Fluorescence Detector

CAUSE

The detector was normalized in a high concentration of organic mobile phase.

The Raman gain was >1000 after the Emission unit normalization.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.