Saturated peaks on FLR detector - WKB94394

Article number: 94394

SYMPTOMS

  • Saturated peaks on the chromatogram

ENVIRONMENT

  • ACQUITY UPLC FLR Detector
  • 2475 Multi-Wavelength Fluorescence Detector

CAUSE

The detector was normalized in a high concentration of organic mobile phase.

The Raman gain was >1000 after the Emission unit normalization.

FIX or WORKAROUND

Repeat the normalization process in 100% water.

ADDITIONAL INFORMATION

 