How to perform the ACQUITY UPLC FLR detector emission units normalization test - WKB238324
Article number: 238324
OBJECTIVE or GOAL
Perform the FLR detector emission units normalization test.
ENVIRONMENT
- ACQUITY UPLC FLR Detector
PROCEDURE
To perform the emission units normalization test
Requirements: Ensure that the FLR detector is powered-on and that you have allowed a 30-minute equilibration period after powering-on.
- After successfully performing the wavelength calibration procedure, ensure that 90:10 water/acetonitrile is flowing through the flow cell at 0.5 mL/min.
- In the ACQUITY console, select FLR Detector from the system tree.
- In the ACQUITY console UPLC FLR detector information window, select Maintain > Normalize Units.
- In the Normalize Units dialog box, click Start.
Result: After the emission units are normalized, the results pane appears.
Requirement: To pass the normalization, the Raman wavelength must be 416 +/- 3 nm.
If unsuccessful, power cycle the detector, clean the flow cell, and repeat the test.
ADDITIONAL INFORMATION
