Article number: 238322

SYMPTOMS

  • Instrument failure
  • "Bad injection volume for partial loop" error appears when making an injection

ENVIRONMENT

  • 2707 Sample Manager
  • Empower
  • Breeze

CAUSE

Loop size and max volume do not match. 

FIX or WORKAROUND

  1.  Update the injection mode based on the loop size in the instrument method.
  2.  See WKB12788 for what injection mode to choose for the loop size.

ADDITIONAL INFORMATION

 