How to perform the wavelength calibration test on an ACQUITY UPLC FLR detector - WKB238326
Article number: 238326
OBJECTIVE or GOAL
Perform the wavelength calibration test on an ACQUITY UPLC FLR detector.
ENVIRONMENT
- ACQUITY UPLC FLR detector
PROCEDURE
To perform the wavelength calibration test
Requirements: Ensure that the FLR detector is powered-on and that you have allowed a 30-minute equilibration period after powering-on.
- Ensure that 90:10 water/acetonitrile is flowing through the flow cell at 0.5 mL/min.
- In the ACQUITY console, select FLR Detector from the system tree.
- In the ACQUITY console UPLC FLR detector information window, select Maintain > Calibrate Wavelength.
- In the Calibrate Wavelength dialog box, click Start.
- After the wavelength calibrates, the Results pane displays any deviations since the last calibration.
Requirement: To pass the calibration, the displayed deviation must be <= 3 nm.
6. If the displayed deviation is not <= 3 nm, repeat the wavelength calibration.
If unsuccessful, power cycle the detector, clean the flow cell, and repeat the test.
ADDITIONAL INFORMATION
id238326, A-30FL, UPFLRARC, UPFLRDET, UPPFLR