Column inlet leaks on ACQUITY UPLC - WKB238414

Article number: 238414

SYMPTOMS

  • Column inlet leaks
  • Shifting Retention times

ENVIRONMENT

  • ACQUITY UPLC (all variants)

CAUSE

Improper column fitment

FIX or WORKAROUND

  1. Remove the column.
  2. Release/loosen all retention screws that may be present.
  3. Ensure that ferrule is in place and that sufficient tube protrudes from its tip.
  4. Fit column and tighten fittings (finger-tight is sufficient).
  5. If the leak persists, replace the ferrule