"WinHttpReceiveResponse failed: 12002" error during upgrading and reloading firmware in ACQUITY modules - WKB238507

Article number: 238507

SYMPTOMS

  • Error during firmware upgrading from a previous version of 1.65 to 1.71 on BSM and reloading 1.65 on SM 
  • Both "Loader" and "Autoloader" are used but do not work
  • CPU appears as CPU2000 Boot Loader

ENVIRONMENT

  • ACQUITY UPLC BSM
  • ACQUITY UPLC SM

CAUSE

Bad network or communication condition

FIX or WORKAROUND

  1. Reseat the network cable, either in the same port or a different port on the network switch.
  2. Open Loader.exe, Scan for instrument and load the firmware. 
  3. If step 2 fails again, replace the Ethernet cable.

ADDITIONAL INFORMATION

 