Low system pressure on ACQUITY BSM - WKB238521

Article number: 238521

SYMPTOMS

  • Low system pressure
  • Only 500 psi is achieved with column attached
  • Accumulator dynamic leak test failure

ENVIRONMENT

  • ACQUITY UPLC BSM

CAUSE

Worn seals or clogged check valves

FIX or WORKAROUND

Rebuild the pump accumulator pump head using part 700004746.

ADDITIONAL INFORMATION