In Empower 3, what are the values behind “Auto” in the Advanced section of the “General” tab in the Instrument Method for ACQUITY UPLC H-Class SM-FTN? - WKB93579
Article number: 93579
ENVIRONMENT
- Empower 3
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class PLUS
ANSWER
The values for "Auto" in the Advanced section of H-Class FTN are as follows:
■ Syringe Draw Rate
- 15 μL needle (standard): 30μL/min
- 30 μL needle: 120 μL/min
■ Needle Placement (from bottom)
For the ANSI-48Vial2mlHolder, the needle tip is lowered to 4 mm from the bottom, and 2 mm from the bottom when using other plates.
■ Air Gap
0 μL for both pre and post
■ Mix Cycles/Mix Stroke Volume
0 for both