Reproducibility is poor on ACQUITY APC system - WKB93601
Article number: 93601
SYMPTOMS
- Baseline not stable on RI trace
- RT change
ENVIRONMENT
- ACQUITY APC
CAUSE
p-ISM fails dynamic leak test. Leak found on primary and/or accumulator pump.
FIX or WORKAROUND
- Flush system with solvent to clear out current solvent. (Use organic solvent to remove organic, water to remove buffer solvents.)
- Wet prime the pump with MeOH.
- Rebuild the pump and replace seals and check valves. Replace any damaged seal wash housing or pump head.
- Wet prime the system with MeOH.
- Put pressure (over 10,000 psi) on new seals for 15 minutes.
- Run the dynamic leak test.
- Run the needle readiness test.
- Run the dynamic leak test. Endpoint column.
ADDITIONAL INFORMATION
id93601, checkvalve, DLT, methanol, UPASMFTN