Reproducibility is poor on ACQUITY APC system - WKB9071
Article number: 9071
SYMPTOMS
- Poor reproducibility
- No leaks or bubbles found
ENVIRONMENT
- ACQUITY APC (FTN)
- ACQUITY H-Class/H-Class (FTN)
- ACQUITY UPLC
- ACQUITY Sample Manager
CAUSE
System startup procedures not performed.
FIX or WORKAROUND
- Prime the solvent lines.
- Prime the washes.
- Characterize the needle seal, needle, and sample loop (Sample Manager) or characterize the needle seal (FTN).
- Equilibrate using the initial conditions until the pressure is stable.
ADDITIONAL INFORMATION
Using the System Startup in the console automates this process. See KCS article # 3505.