What cables are connected to the Waters Fraction Collector III when controlled by Empower software? - WKB9065
Article number: 9065
ENVIRONMENT
- Waters Fraction Collector III
    - Part 205000823 for the upgrade kit, WFC III w/Empower v1.10
 
- 
    Empower 1154 SP L
- 
    Empower 2154 FR 5
- Empower 3
ANSWER
When connecting the WFC III for operation with Empower you must attach the following cables:
- Power cord
- Contact closure cable for a start signal [WFC3 > Injector's Inject Start (for e.g. Alliance)]
- Analog cable for the analog signal in [WFC3 > Detector's Analog Out (remember to configure the corresponding instrument method as well)]
- RS-232 cable (PC COM Port 1 or 2)
    - Note: Baud rate settings are automatically applied by the ICS software.
- If a COM port is not available order 668000407 - 1 Edgeport USB-TO-SERIAL Converter
 
ADDITIONAL INFORMATION

Contact closure/remote start cable is WAT062031 (CHART MARKER CABLE ASSY).
RS232 cable is 700002170 (RS-232 SERIAL CABLE WITH RFC CORE-8).