What are normal pressures for a nanoACQUITY or ACQUITY M-class in single pump trapping with a 100 um column? - WKB9073
Article number: 9073
ENVIRONMENT
- nanoACQUITY
- ACQUITY UPLC M-class
- Single pump trapping mode
- 100um Analytical column
- typical nanoACQUITY Pressure
ANSWER
%A | %B |
Flow Rate (uL/min) |
Pressure | Trapping |
97 | 3 | 0.4 | 2924 | No |
50 | 50 | 0.4 | 2800 | No |
15 | 85 | 0.4 | 1602 | No |
10 | 90 | 0.4 | 1412 | No |
99 | 1 | 5 | 1232 | yes |
99 | 1 | 15 | 3720 | yes |
97 | 3 | 15 | 3785 | yes |
50 | 50 | 15 | 3192 | yes |
15 | 85 | 15 | 1704 | yes |
97 | 3 | 0.3 | 2400 | No |
50 | 50 | 0.3 | 2040 | No |
15 | 85 | 0.3 | 1161 | No |
10 | 90 | 0.3 | 1062 | No |
ADDITIONAL INFORMATION
id9073, UPBINARY, UPMASM, UPMBSM, UPMSMFL, UPNAYXPMP, UPNSAMMGR, UPNSOLMGR