Sample Fluidics High Pressure Limit (value) on ACQUITY UPLC Sample Manager- WKB67490

Article number: 67490

SYMPTOMS

  • Sample Fluidics High Pressure Limit (value) error
  • Error is triggered during wash syringes priming 
  • Wash Syringe move h/w fault occurs in combination with the main one
  • Errors occur randomly 

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC I-Class

CAUSE

Blocked injection valve cartridge.

FIX or WORKAROUND

  1. Replace the injection valve cartridge.

ADDITIONAL INFORMATION

UPLC Sample Manager injection valve Part Number 700002765.

H Class injection valve Part Number 700005236

H Class Bio injection valve Part Number 700005407

I Class FTN injection valve Part Number 700006057

I Class FL injection valve Part Number 700006069

Prior to replace the injection valve cartridge, try to remove the clog according to WKB9960. 

For troubleshooting see also WKB19544.