ACQUITY TUV/PDA "Filter initialization failure" - WKB67475
Article number: 67475
SYMPTOMS
- Filter initialization failure during wavelength calibration
- Grating initialization failure during start-up
- High backpressure isolated to the flowcell
ENVIRONMENT
- Empower
- ACQUITY UPLC PDA Detector
- ACQUITY UPLC Tunable UV Detector
- 2996 PDA Detector
- 2998 Photodiode Array Detector
- 2489 UV/Vis Detector
- 2487 Dual-wavelength Absorbance Detector
- 2475 Multi-Wavelength Fluorescence Detector
- ACQUITY UPLC FLR Detector
CAUSE
Dirty, contaminated, or defective flow cell
FIX or WORKAROUND
- Flush the flow cell with aqueous solvent (back flushing at low flow rate may be necessary)
- After aqueous wash, clean flow cell with 25:25:25:25, Water:Methanol:Acetonitrile:Isopropyl alcohol
- Verify functionality or replace flow cell as a billable consumable
ADDITIONAL INFORMATION
Part numbers for TUV Flowcell
- ACQUITY Cell TUV, Standard (# 205015000)
- ACQUITY UPLC ID Cell TUV, Analytical - (# 205015016)
- ACQUITY UPLC TUV Flow Cell, Low Flow (# 205015001)
- ACQUITY UPLC ID Cell TUV, HS FC (# 205015018)
- ACQUITY UPLC TUV 5 mm SS FC (# 205000610)
- ACQUITY TUV, Flow Cell, 10mm, 250 nL (# 205015032)
- ACQUITY TUV, Flow Cell, 10mm 500nL (# 205015033)
- ACQUITY UPLC TUV 5 mm Titanium FC (# 205000611)
Part numbers for PDA Flowcell
- ACQUITY PDA, Flowcell, 10mm 500nL (# 205015036)
- ACQUITY PDA, Flowcell, 10mm 250nL (# 205015035)
- ACQUITY PDA, Flowcell, 5mm ss (#205000612)
- High-sensitivity Flow Cell (25-mm) (# 205015019)
- High-throughput Flow Cell (3-mm) (# 205015006)