Linearity height sensitivity failure of SQT 2695 with UV detector - WKB65210

Article number: 65210

SYMPTOMS

  • Detector linearity height sensitivity failure
  • System readiness check failure for detector response

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2998 Photodiode Array Detector
  • 2489 UV/Vis Detector
  • SQT HPLC Version K
  • SQT HPLC Version P

CAUSE

Bad connection/tubing on the column

Poor recovery and failure of the height RSD results

FIX or WORKAROUND

Replace the tubing from seal pack to column and column to detector.

430000922 - 0.005" tubing from injector to column to detector 

ADDITIONAL INFORMATION

What is the part number and ID of tubing for the red line coming from 2695 injector going into the 2489 detector? - WKB8847

id65210, 2489, 2489B, 2489SFC, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, 2998, 2998B, 2998SFC, A-10PDA, A-10SSM, A-10UV, A-30PDA, A-30UV, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, EMP2QLIC, EMP3Q1ULIC, EMP3QLIC, UPAPDA, UPARPTUV, UPC2PDA, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA, UPPTUV, UPTUV, UPTUVARC, UPTUVARCB, UPTUV-E, UPTUVTC