No flow from ACQUITY UPLC solvent manager when priming - WKB65252
Article number: 65252
SYMPTOMS
- No flow from one or both pumps
- This can happen intermittently; sometimes one pump has a flow when primed and even passes the dynamic leak test, but the system pressure drops to zero after 5 to 15 minutes of normal operation
- Pressure trace indicates close to zero pressure for the faulty pump's primary
ENVIRONMENT
- ACQUITY UPLC BSM
- ACQUITY UPLC H-Class QSM/Bio QSM
- ACQUITY UPLC I-Class BSM
- ACQUITY UPLC H-Class+ QSM
- ACQUITY UPLC ISM
CAUSE
Faulty check valves
FIX or WORKAROUND
Replace the check valves on the affected pump.
ACQUITY BSM Assy CV Cartridge i2V, 700003563
ACQUITY BSM Assy CV Primary Spring-loaded, 700002596
ACQUITY BSM Assy CV Accumulator Dual ball and seat, 700002968
ACQUITY QSM Assy CV Cartridge i2V, 700005165
ACQUITY QSM Assy CV Accumulator Dual ball and seat, 700005164
ACQUITY QSM Assy CV Accumulator TI, 700005414
ADDITIONAL INFORMATION
Run the dynamic leak test.
id65252, A-30SDM, checkvalve, DLT, UPAQSM, UPBINARY, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM, UPPQSM, UPQSM, UPQSM+, UPQSMBIO