Skip to main content
Waters

No flow from ACQUITY UPLC solvent manager when priming - WKB65252

Article number: 65252

SYMPTOMS

  • No flow from one or both pumps
  • This can happen intermittently; sometimes one pump has a flow when primed and even passes the dynamic leak test, but the system pressure drops to zero after 5 to 15 minutes of normal operation
  • Pressure trace indicates close to zero pressure for the faulty pump's primary

ENVIRONMENT

  • ACQUITY UPLC BSM
  • ACQUITY UPLC H-Class QSM/Bio QSM
  • ACQUITY UPLC I-Class BSM
  • ACQUITY UPLC H-Class+ QSM
  • ACQUITY UPLC ISM

CAUSE

Faulty check valves

FIX or WORKAROUND

Replace the check valves on the affected pump.

ACQUITY BSM Assy CV Cartridge i2V,  700003563

ACQUITY BSM Assy CV Primary Spring-loaded, 700002596

ACQUITY BSM Assy CV Accumulator Dual ball and seat, 700002968

ACQUITY QSM Assy CV Cartridge i2V,  700005165

ACQUITY QSM Assy CV Accumulator Dual ball and seat, 700005164

ACQUITY QSM Assy CV Accumulator TI, 700005414

ADDITIONAL INFORMATION

Run the dynamic leak test.

Not able to find a solution? Click here to request help.