No QDa signal after diverter valve is fitted - WKB65159
Article number: 65159
SYMPTOMS
- QDa shows no signal
- UV detector operates as expected
- Diverter valve was recently fitted to QDa
ENVIRONMENT
- ACQUITY QDa Mass Detector
CAUSE
The diverter valve is set to waste (position 2).
FIX or WORKAROUND
Update the instrument methods to include an event in the QDa Method Events Editor.
ADDITIONAL INFORMATION
See page 2 of the ACQUITY Diverter Valve Quick Reference Card (715005336).
id65159, QDA