No QDa signal after diverter valve is fitted - WKB65159

Article number: 65159

SYMPTOMS

  • QDa shows no signal
  • UV detector operates as expected
  • Diverter valve was recently fitted to QDa

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

The diverter valve is set to waste (position 2).

FIX or WORKAROUND

Update the instrument methods to include an event in the QDa Method Events Editor.

ADDITIONAL INFORMATION

See page 2 of the ACQUITY Diverter Valve Quick Reference Card (715005336).