ACQUITY UPLC Tunable UV Detector flow cell leaks - WKB56055
Article number: 56055
SYMPTOMS
- The peak of interest does not appear, and the chromatographic baseline is noisy
- Corrosion is visible around the lower flow cell body
- The drip tray has buildup of solvent precipitate
- The back pressure regulator has salt buildup on the edges
- Pulling the flow cell out and running flow reveals leaks on the flow cell tubing
ENVIRONMENT
- H-Class
- ACQUITY UPLC Tunable UV Detector
CAUSE
Damaged / defective flow cell (possibly due to high salts and various solvent buildup causing high back pressure)
FIX or WORKAROUND
- Replace the flow cell (see WKB61002)
- Verify functionality by performing initialization and wavelength calibration successfully
ADDITIONAL INFORMATION
In some cases, if the salt buildup is after the flow cell, it may be advisable to replace the back pressure regulator (700002676).
For more flow cell options/part numbers, see article What are the UV cell options and wetted materials? - WKB80653
id56055, A-30UV, UPARPTUV, UPBINARY, UPPTUV, UPTUV, UPTUVARC, UPTUVARCB, UPTUV-E, UPTUVTC