Skip to main content
Waters

HDX salt buildup around syringe shield - WKB56290

Article number: 56290

SYMPTOMS

Salt is building up around the sensor shield on the HDX autosampler arm during sample acquisitions.

ENVIRONMENT

HDX

CAUSE

Syringe plunger seal is damaged and allowing solvent past the plunger to sit in the top of the syringe.

FIX or WORKAROUND

Replace affected syringe.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.